Specifications
  CA 3000, Automatic CM 3000, Manual
Positive Pressure Settable from 5 to 60 cm of H2O Settable from 5 to 60 cm of H2O
Negative Pressure Settable from 5 to 60 cm of H2O Settable from 5 to 60 cm of H2O
Typical Inhalation Flow 3.3 liters/sec. when set to minimum; equals exhalation flow when set to maximum 3.3 liters/sec. when set to minimum; equals exhalation flow when set to maximum
Typical Exhalation Flow 10 liters/sec. (actual flow depends on set pressure and patient airway resistance) 10 liters/sec. (actual flow depends on set pressure and patient airway resistance)
Pressure Gauge -70 to 0 to +70 cm H2O
(accuracy 6 cm H2O)
-70 to 0 to +70 cm H2O
(accuracy 6 cm H2O)
Mode of Operation Automatic and manual timing Manual timing
Inhalation, Exhalation and Pause Times 0-5 sec. (in automatic mode) or user variable (in manual mode) User variable
Blower Type Two-stage centrifugal blower with AC/DC universal motor Two-stage centrifugal blower with AC/DC universal motor
Input Voltage 100-120VAC, 60 Hz 100-120VAC, 60 Hz
Input Power 300 VA 300 VA
Dimensions
(H x W x D)
292 mm x 279 mm x 419 mm
(11.5 x 11 x 16.5 in)
292 mm x 279 mm x 419 mm
(11.5 x 11 x 16.5 in)
Weight 11 kg (24 lbs) 9.3 kg (20.6 lbs.)

Top Of Page